Calculate the viscosity from a Boekel Zahn Cup Viscosimeter using the cup size and efflux time.
The viscosity is measured by determining the amount of time it takes for a steady stream of sample to flow from the orifice in the bottom of the cup. This is correlated to the viscosity in centistokes. All measurements should be made at 77 F (25 C).
Cup Capacity - 44ml
Cup #1
|
Cup #2
|
Cup #3
|
Cup #4
|
Cup #5
|
|
Model No. 27134 |
-001
|
-002
|
-003
|
-004
|
-005
|
Orifice Diameter (in) |
0.078"
|
0.108"
|
0.148"
|
0.168"
|
0.208"
|
Zahn Range (sec) |
45-80
|
25-80
|
20-75
|
20-80
|
20-75
|
Centistoke Range |
18-56
|
40-230
|
150-790
|
220-1100
|
460-1725+
|
Applications |
Very Thin Oil
|
Thin Oil or
Lacquer |
Medium Oil or
Mixed Paints |
Heavy Mixtures
|
Very Heavy
Mixtures |
See ASTM #D 4212: Viscosity by dip-type viscosity cups.
Debut: October 25, 2003. Revision No. 2.
Please send comments on this page to Jeffrey Clymer.